Scope Camera MVI Nikon SMZ 1000

Edit this page

Installing and configuring MVI Scope Camera.


Install

  • Download software from Google Drive folder (Lotterhos Lab > equipment > MVI Nikon).

  • Run as Administrator (PC) to install.

note - a message might pop up saying the key is missing; ignore and continue installing without the key

  • Under NIS-Elements Package, select ‘Local Option’

  • Under Cameras, select Nikon Fi3

  • Under Modules, select EDF, Local options > Language Pack

  • Under Devices, select Manual Microscope and Nikon SMZ manual

Calibration settings

  • Plug dongle and camera to laptop.

At start up: first power up camera, second run application software (reverse to shut down).

  • Acquire > Select driver > Nikon DS Fi3

  • Acquire Camera light path > relay lens = 0.70 x > ok

  • View > Acquisition controls > SMZ Pad > Nosepiece > Plan Apo1x

note - for zoom calibration, double check the the value on the software matches what you’re using on the scope (knob on the right side of the ocular) and update the software each time you use a different zoom before taking a photo (zoom cal can not be corrected after an image has been captured).